Geoprobe Systems® developed the GW1600 Pneumatic Slug Test Kit for use with Geoprobe® brand groundwater sampling, profiling, and monitoring tools to conduct pneumatic slug tests in saturated formations of unconsolidated soils or sediments. The hydraulic conductivity test system provides high-quality data for accurate determination of formation hydraulic conductivity (K). This is an important parameter in contaminant investigations as hydraulic conductivity is useful in determining the volume of water passing through a given soil cross section, and for estimating contaminant flux. This information can then be used to determine potential “risk” as well as guiding the design or implementation of remedial systems, whether that be injection of fluids, pump and treat or other remedial methods.
= Tool String Diagram (TSD)
Geoprobe Systems® developed the GW1600 Pneumatic Slug Test Kit for use with Geoprobe® brand groundwater sampling, profiling, and monitoring tools to conduct pneumatic slug tests in saturated formations of unconsolidated soils or sediments. Adapters also permit use of the PST system on conventional PVC monitoring wells. The hydraulic conductivity test system provides high-quality data for accurate determination of formation hydraulic conductivity (K). The slug test kit is self-contained and is easily taken from location to location. In pneumatic slug testing, the well head is sealed and air pressure or vacuum is used to change the static water level. As air pressure in the well is increased or decreased, the water level changes until the water pressure "up" and the air pressure "down" are equal. Once the water level is stable, a release valve is quickly opened, instantaneously releasing the air pressure or vacuum. The water level recovers without splashing and the pressure transducer and data logger/computer record the changes in water level and time (Figure 2).
The hydraulic conductivity test kit includes the necessary adapters and manifold (Figure 1) you will need for slug testing in screens installed with Geoprobe® 1.25, 1.5, 1.75 in. rods. Adapters for use with 0.5in. - 2.0in. (12.7 to 50.8mm) diameter PVC monitoring wells are also available. Hydraulic conductivity (K) is an important parameter in contaminant investigations. Hydraulic conductivity (K) is useful in determining the volume of water passing through a given soil cross section, and for estimating contaminant flux. With the GW1600 Pneumatic Slug Test Kit, set-up is simple, requires only a few minutes, and all of the hydraulic conductivity test equipment needed is easily transported from one location to another.
Field Setup and Slug Test for Groundwater Sampler or Direct Push Monitoring Well
Figure 4 shows what is happening in the well during a rising head slug test.
This figure shows data obtained in a moderate K formation using a pneumatic slug test technique. This hydraulic conductivity test data, taken in glacial soil in northern Illinois, shows a single test cycle of pressurization followed by a release of air pressure and recovery (formation response).
Assembly and Driving
The SP16 sampler is assembled with fresh O-rings to provide a sealed system while advancing the tool string. An O-ring kit is available for these tools and details of assembly are provided in the standard operating procedure (link below). After each rod is driven, an O-ring is placed on the rod before the next rod is added. All rod joints must be sealed to prevent air leaks and allow for pneumatic slug testing. Additional probe rods are added to the tool string and driven until the end of the screen is at the desired depth.
Deployment
Once the screen is at the desired depth, extension rods with a screen push adapter (GW1535) are inserted through the bore of the rods. While holding down on the extension rod handle, the probe rods and screen sheath are retracted with the rod grip system to expose the screen to the formation. Knowing the exact length of screen exposed to the formation is critical for calculating the hydraulic conductivity (K). Determine beforehand the length of screen to be exposed and mark this on the extension rod above the top probe rod. Be sure to add 2 inches to the screen length to account for the shaft of the expendable drive point and blank section at base of screen (Figure 4).
Screen Development
Development is a key step required to make an accurate determination of the hydraulic conductivity (K) of the screened formation. Development is needed to reestablish natural flow from the formation into the screen (ASTM 5521). Even in “clean” sand aquifers, conducting a slug test before development routinely yield determinations of hydraulic conductivity (K) 3 to 5 times less than slug tests after development (Butler et al. 2002). Development of the SP16 in sandy formations may be completed using the tubing check valve system (MN 214061). Oscillating and purging through the screen interval with the check valve is effective.
Slug Testing Fine Grained Formations
So you need to conduct a slug test in a fine-grained formation. Yet you know driving the SP16 groundwater samplers or advancing the DT22 rod system into the formation will cause compression and bias your hydraulic conductivity test results. Your best option... drive the 2.25 in. dual tube rods down to the top of the interval you want to slug test. Retract the inner 1 in. rods and attach the Soil Precore Drive head and sample tube. Lower the sampler to depth and core the soil below the SP22 Profiler screen. You may want to "brush" the cored formation before installing the screen with the BU600 brush to relieve smearing. Now you are ready to slug test in the fine grained low-K formation. A simple falling head test is most often used under these conditions.
Hydraulic Conductivity (K) is a measure of the capability of a medium to transmit water.
This is an important parameter in contaminant investigations as hydraulic conductivity is useful in determining the volume of water passing through a given soil cross section, and for estimating the flow velocity of contaminated groundwater.
In the past, hydraulic conductivity was typically measured using pumping tests or by performing “slug tests” of installed wells. Both of these past hydraulic conductivity test practices, while having their technical strengths, required the installation of some form of well casing; either of the pumping or monitoring type.
What is a slug test?
As defined by Fetter (1994) a slug test is “an aquifer test made by pouring a small instantaneous charge of water into a well or by withdrawing a slug of water from the well” (e.g with a bailer). This definition clearly shows how early slug testing was conducted, by quickly adding water to a well or quickly removing a bailer full of water from the well.
When environmental monitoring wells are being tested, you don’t want to add (or remove) water from the well. This either alters the ambient water quality or generates potentially hazardous waste. It quickly became common for people to use a “mechanical slug” to run a slug test. This is simply a 3 to 5 foot long section of PVC pipe filled with sand and capped on each end. A cord is attached to one end of the mechanical slug so it can be quickly lowered below the water level (slug-in test/falling head test) and then later quickly raised above the water level (slug-out test/rising head test). Again the change in water level and time are recorded.
Quickly lowering or raising the mechanical slug in or out of the water may cause splashing in the well. For slug tests lasting several minutes or longer this is not a significant problem. However, for wells that recover from a slug test in less than a minute the splashing will interfere with a lot of the early time data so that a good determination of the changes in the water level can not be made. Because of these problems the pneumatic slug testing method was developed (Prosser 1981, Butler 1997). In this method, the well head is sealed and air pressure is used to displace the water level.
Why calculate the hydraulic conductivity of a formation?
Since the hydraulic conductivity lets you know how fast the groundwater can move through a formation, it is the first step in knowing how fast contaminants may be able to move. This information is used to determine the potential “risk” caused by the presence and migration of contaminants in the subsurface for risk based corrective actions (RBCA). It is also a vital piece of information required to determine if monitored natural attenuation (MNA) is an acceptable remedial option for a contaminated facility (EPA 1998). Additionally, knowing the hydraulic conductivity is a very useful piece of information when implementing or designing a remedial system, be it injection of fluids, pump and treat, or other remedial methods.
Why use Direct Push methods to conduct slug tests?
There are several reasons, but primarily economics is the driver. You can save thousands of dollars and a lot of time using direct push methods to do this work. A summary of specific reasons include:
Can direct push slug tests provide results comparable to tests conducted in conventional monitoring wells?
In short, YES. When the screens are installed properly and adequately developed, field comparison has shown that the direct push methods provide essentially the same results as conventional monitoring wells screened over the same interval (Butler et al. 2002, McCall et al., 2002).
Where did the slug come from?
Slug Test. No, it’s not a test to see if you have slimy worms crawling around in your garden! When single well hydraulic conductivity tests first got started, the wells being tested were pretty large in diameter ... 4-in., 8-in. or even larger. There were no neat small pressure transducers, no data loggers, and definitely no portable computers at this time. The tests were completed by pouring a 5-gallon bucket of water down the well casing and measuring the change in water level with a tape, marking time using your wrist watch, and writing down the time and water level in your field book with your No. 2 lead pencil! The 5-gallons of water was simply a big “slug” of water poured down the well. And presto! The “Slug Test” was born.
Features & Options
Geoprobe Systems® developed the GW1600 Pneumatic Slug Test Kit for use with Geoprobe® brand groundwater sampling, profiling, and monitoring tools to conduct pneumatic slug tests in saturated formations of unconsolidated soils or sediments. Adapters also permit use of the PST system on conventional PVC monitoring wells. The hydraulic conductivity test system provides high-quality data for accurate determination of formation hydraulic conductivity (K). The slug test kit is self-contained and is easily taken from location to location. In pneumatic slug testing, the well head is sealed and air pressure or vacuum is used to change the static water level. As air pressure in the well is increased or decreased, the water level changes until the water pressure "up" and the air pressure "down" are equal. Once the water level is stable, a release valve is quickly opened, instantaneously releasing the air pressure or vacuum. The water level recovers without splashing and the pressure transducer and data logger/computer record the changes in water level and time (Figure 2).
The hydraulic conductivity test kit includes the necessary adapters and manifold (Figure 1) you will need for slug testing in screens installed with Geoprobe® 1.25, 1.5, 1.75 in. rods. Adapters for use with 0.5in. - 2.0in. (12.7 to 50.8mm) diameter PVC monitoring wells are also available. Hydraulic conductivity (K) is an important parameter in contaminant investigations. Hydraulic conductivity (K) is useful in determining the volume of water passing through a given soil cross section, and for estimating contaminant flux. With the GW1600 Pneumatic Slug Test Kit, set-up is simple, requires only a few minutes, and all of the hydraulic conductivity test equipment needed is easily transported from one location to another.
Field Setup and Slug Test for Groundwater Sampler or Direct Push Monitoring Well
Figure 4 shows what is happening in the well during a rising head slug test.
This figure shows data obtained in a moderate K formation using a pneumatic slug test technique. This hydraulic conductivity test data, taken in glacial soil in northern Illinois, shows a single test cycle of pressurization followed by a release of air pressure and recovery (formation response).
Assembly and Driving
The SP16 sampler is assembled with fresh O-rings to provide a sealed system while advancing the tool string. An O-ring kit is available for these tools and details of assembly are provided in the standard operating procedure (link below). After each rod is driven, an O-ring is placed on the rod before the next rod is added. All rod joints must be sealed to prevent air leaks and allow for pneumatic slug testing. Additional probe rods are added to the tool string and driven until the end of the screen is at the desired depth.
Deployment
Once the screen is at the desired depth, extension rods with a screen push adapter (GW1535) are inserted through the bore of the rods. While holding down on the extension rod handle, the probe rods and screen sheath are retracted with the rod grip system to expose the screen to the formation. Knowing the exact length of screen exposed to the formation is critical for calculating the hydraulic conductivity (K). Determine beforehand the length of screen to be exposed and mark this on the extension rod above the top probe rod. Be sure to add 2 inches to the screen length to account for the shaft of the expendable drive point and blank section at base of screen (Figure 4).
Screen Development
Development is a key step required to make an accurate determination of the hydraulic conductivity (K) of the screened formation. Development is needed to reestablish natural flow from the formation into the screen (ASTM 5521). Even in “clean” sand aquifers, conducting a slug test before development routinely yield determinations of hydraulic conductivity (K) 3 to 5 times less than slug tests after development (Butler et al. 2002). Development of the SP16 in sandy formations may be completed using the tubing check valve system (MN 214061). Oscillating and purging through the screen interval with the check valve is effective.
Slug Testing Fine Grained Formations
So you need to conduct a slug test in a fine-grained formation. Yet you know driving the SP16 groundwater samplers or advancing the DT22 rod system into the formation will cause compression and bias your hydraulic conductivity test results. Your best option... drive the 2.25 in. dual tube rods down to the top of the interval you want to slug test. Retract the inner 1 in. rods and attach the Soil Precore Drive head and sample tube. Lower the sampler to depth and core the soil below the SP22 Profiler screen. You may want to "brush" the cored formation before installing the screen with the BU600 brush to relieve smearing. Now you are ready to slug test in the fine grained low-K formation. A simple falling head test is most often used under these conditions.
Hydraulic Conductivity (K) is a measure of the capability of a medium to transmit water.
This is an important parameter in contaminant investigations as hydraulic conductivity is useful in determining the volume of water passing through a given soil cross section, and for estimating the flow velocity of contaminated groundwater.
In the past, hydraulic conductivity was typically measured using pumping tests or by performing “slug tests” of installed wells. Both of these past hydraulic conductivity test practices, while having their technical strengths, required the installation of some form of well casing; either of the pumping or monitoring type.
What is a slug test?
As defined by Fetter (1994) a slug test is “an aquifer test made by pouring a small instantaneous charge of water into a well or by withdrawing a slug of water from the well” (e.g with a bailer). This definition clearly shows how early slug testing was conducted, by quickly adding water to a well or quickly removing a bailer full of water from the well.
When environmental monitoring wells are being tested, you don’t want to add (or remove) water from the well. This either alters the ambient water quality or generates potentially hazardous waste. It quickly became common for people to use a “mechanical slug” to run a slug test. This is simply a 3 to 5 foot long section of PVC pipe filled with sand and capped on each end. A cord is attached to one end of the mechanical slug so it can be quickly lowered below the water level (slug-in test/falling head test) and then later quickly raised above the water level (slug-out test/rising head test). Again the change in water level and time are recorded.
Quickly lowering or raising the mechanical slug in or out of the water may cause splashing in the well. For slug tests lasting several minutes or longer this is not a significant problem. However, for wells that recover from a slug test in less than a minute the splashing will interfere with a lot of the early time data so that a good determination of the changes in the water level can not be made. Because of these problems the pneumatic slug testing method was developed (Prosser 1981, Butler 1997). In this method, the well head is sealed and air pressure is used to displace the water level.
Why calculate the hydraulic conductivity of a formation?
Since the hydraulic conductivity lets you know how fast the groundwater can move through a formation, it is the first step in knowing how fast contaminants may be able to move. This information is used to determine the potential “risk” caused by the presence and migration of contaminants in the subsurface for risk based corrective actions (RBCA). It is also a vital piece of information required to determine if monitored natural attenuation (MNA) is an acceptable remedial option for a contaminated facility (EPA 1998). Additionally, knowing the hydraulic conductivity is a very useful piece of information when implementing or designing a remedial system, be it injection of fluids, pump and treat, or other remedial methods.
Why use Direct Push methods to conduct slug tests?
There are several reasons, but primarily economics is the driver. You can save thousands of dollars and a lot of time using direct push methods to do this work. A summary of specific reasons include:
Can direct push slug tests provide results comparable to tests conducted in conventional monitoring wells?
In short, YES. When the screens are installed properly and adequately developed, field comparison has shown that the direct push methods provide essentially the same results as conventional monitoring wells screened over the same interval (Butler et al. 2002, McCall et al., 2002).
Where did the slug come from?
Slug Test. No, it’s not a test to see if you have slimy worms crawling around in your garden! When single well hydraulic conductivity tests first got started, the wells being tested were pretty large in diameter ... 4-in., 8-in. or even larger. There were no neat small pressure transducers, no data loggers, and definitely no portable computers at this time. The tests were completed by pouring a 5-gallon bucket of water down the well casing and measuring the change in water level with a tape, marking time using your wrist watch, and writing down the time and water level in your field book with your No. 2 lead pencil! The 5-gallons of water was simply a big “slug” of water poured down the well. And presto! The “Slug Test” was born.
The kit includes everything you will need for slug testing in screens installed with Geoprobe® 1.25 or 1.5 in. rods. Kit includes the pneumatic manifold, gauges, rod adapters, pressure transducer, a hand pump for applying air pressure or vacuum to the well head, data logger that adapts to the user's PC along with the Aquisition software.
With the Pneumatic Slug Test Kit (MN 214038), set-up is simple, requires only a few minutes, all of the essential assemblies and accessories are included to prepare you for measuring hydraulic conductivity with Geoprobe® groundwater samplers in the field. The kit is self contained and is easy to move from location to location. Optional PVC adapters allow for pneumatic slug testing on PVC monitoring wells from 0.5in. to 2.0in. nominal diameter.
Other features of the system include:
Click on a section below to view information.
Since 1987, Geoprobe® has manufactured innovative drilling rigs and tooling - engineered for efficiency and safety - simplifying drillers’ jobs and empowering their companies to succeed as productive and profitable leaders in the industry. When you partner with Geoprobe® you receive:
Engineering and building industry-leading drilling rigs, tooling, and techniques for the technical driller based on your needs to work safer and more efficiently.
Ensuring drilling rigs and tooling are created in conjunction – with consistent quality – to collect the highest-quality information with the most accurate result to get you to, into, and through the job efficiently.
Equipping you to do your best job and keeping you in the field via one-on-one expert sales and service technicians manning live support phone lines, shipping necessary parts same-day.
Geoprobe Systems® developed the GW1600 Pneumatic Slug Test Kit for use with Geoprobe® brand groundwater sampling, profiling, and monitoring tools to conduct pneumatic slug tests in saturated formations of unconsolidated soils or sediments. Adapters also permit use of the PST system on conventional PVC monitoring wells. The hydraulic conductivity test system provides high-quality data for accurate determination of formation hydraulic conductivity (K). The slug test kit is self-contained and is easily taken from location to location. In pneumatic slug testing, the well head is sealed and air pressure or vacuum is used to change the static water level. As air pressure in the well is increased or decreased, the water level changes until the water pressure "up" and the air pressure "down" are equal. Once the water level is stable, a release valve is quickly opened, instantaneously releasing the air pressure or vacuum. The water level recovers without splashing and the pressure transducer and data logger/computer record the changes in water level and time (Figure 2).
The hydraulic conductivity test kit includes the necessary adapters and manifold (Figure 1) you will need for slug testing in screens installed with Geoprobe® 1.25, 1.5, 1.75 in. rods. Adapters for use with 0.5in. - 2.0in. (12.7 to 50.8mm) diameter PVC monitoring wells are also available. Hydraulic conductivity (K) is an important parameter in contaminant investigations. Hydraulic conductivity (K) is useful in determining the volume of water passing through a given soil cross section, and for estimating contaminant flux. With the GW1600 Pneumatic Slug Test Kit, set-up is simple, requires only a few minutes, and all of the hydraulic conductivity test equipment needed is easily transported from one location to another.
Field Setup and Slug Test for Groundwater Sampler or Direct Push Monitoring Well
Figure 4 shows what is happening in the well during a rising head slug test.
This figure shows data obtained in a moderate K formation using a pneumatic slug test technique. This hydraulic conductivity test data, taken in glacial soil in northern Illinois, shows a single test cycle of pressurization followed by a release of air pressure and recovery (formation response).
Assembly and Driving
The SP16 sampler is assembled with fresh O-rings to provide a sealed system while advancing the tool string. An O-ring kit is available for these tools and details of assembly are provided in the standard operating procedure (link below). After each rod is driven, an O-ring is placed on the rod before the next rod is added. All rod joints must be sealed to prevent air leaks and allow for pneumatic slug testing. Additional probe rods are added to the tool string and driven until the end of the screen is at the desired depth.
Deployment
Once the screen is at the desired depth, extension rods with a screen push adapter (GW1535) are inserted through the bore of the rods. While holding down on the extension rod handle, the probe rods and screen sheath are retracted with the rod grip system to expose the screen to the formation. Knowing the exact length of screen exposed to the formation is critical for calculating the hydraulic conductivity (K). Determine beforehand the length of screen to be exposed and mark this on the extension rod above the top probe rod. Be sure to add 2 inches to the screen length to account for the shaft of the expendable drive point and blank section at base of screen (Figure 4).
Screen Development
Development is a key step required to make an accurate determination of the hydraulic conductivity (K) of the screened formation. Development is needed to reestablish natural flow from the formation into the screen (ASTM 5521). Even in “clean” sand aquifers, conducting a slug test before development routinely yield determinations of hydraulic conductivity (K) 3 to 5 times less than slug tests after development (Butler et al. 2002). Development of the SP16 in sandy formations may be completed using the tubing check valve system (MN 214061). Oscillating and purging through the screen interval with the check valve is effective.
Slug Testing Fine Grained Formations
So you need to conduct a slug test in a fine-grained formation. Yet you know driving the SP16 groundwater samplers or advancing the DT22 rod system into the formation will cause compression and bias your hydraulic conductivity test results. Your best option... drive the 2.25 in. dual tube rods down to the top of the interval you want to slug test. Retract the inner 1 in. rods and attach the Soil Precore Drive head and sample tube. Lower the sampler to depth and core the soil below the SP22 Profiler screen. You may want to "brush" the cored formation before installing the screen with the BU600 brush to relieve smearing. Now you are ready to slug test in the fine grained low-K formation. A simple falling head test is most often used under these conditions.
Hydraulic Conductivity (K) is a measure of the capability of a medium to transmit water.
This is an important parameter in contaminant investigations as hydraulic conductivity is useful in determining the volume of water passing through a given soil cross section, and for estimating the flow velocity of contaminated groundwater.
In the past, hydraulic conductivity was typically measured using pumping tests or by performing “slug tests” of installed wells. Both of these past hydraulic conductivity test practices, while having their technical strengths, required the installation of some form of well casing; either of the pumping or monitoring type.
What is a slug test?
As defined by Fetter (1994) a slug test is “an aquifer test made by pouring a small instantaneous charge of water into a well or by withdrawing a slug of water from the well” (e.g with a bailer). This definition clearly shows how early slug testing was conducted, by quickly adding water to a well or quickly removing a bailer full of water from the well.
When environmental monitoring wells are being tested, you don’t want to add (or remove) water from the well. This either alters the ambient water quality or generates potentially hazardous waste. It quickly became common for people to use a “mechanical slug” to run a slug test. This is simply a 3 to 5 foot long section of PVC pipe filled with sand and capped on each end. A cord is attached to one end of the mechanical slug so it can be quickly lowered below the water level (slug-in test/falling head test) and then later quickly raised above the water level (slug-out test/rising head test). Again the change in water level and time are recorded.
Quickly lowering or raising the mechanical slug in or out of the water may cause splashing in the well. For slug tests lasting several minutes or longer this is not a significant problem. However, for wells that recover from a slug test in less than a minute the splashing will interfere with a lot of the early time data so that a good determination of the changes in the water level can not be made. Because of these problems the pneumatic slug testing method was developed (Prosser 1981, Butler 1997). In this method, the well head is sealed and air pressure is used to displace the water level.
Why calculate the hydraulic conductivity of a formation?
Since the hydraulic conductivity lets you know how fast the groundwater can move through a formation, it is the first step in knowing how fast contaminants may be able to move. This information is used to determine the potential “risk” caused by the presence and migration of contaminants in the subsurface for risk based corrective actions (RBCA). It is also a vital piece of information required to determine if monitored natural attenuation (MNA) is an acceptable remedial option for a contaminated facility (EPA 1998). Additionally, knowing the hydraulic conductivity is a very useful piece of information when implementing or designing a remedial system, be it injection of fluids, pump and treat, or other remedial methods.
Why use Direct Push methods to conduct slug tests?
There are several reasons, but primarily economics is the driver. You can save thousands of dollars and a lot of time using direct push methods to do this work. A summary of specific reasons include:
Can direct push slug tests provide results comparable to tests conducted in conventional monitoring wells?
In short, YES. When the screens are installed properly and adequately developed, field comparison has shown that the direct push methods provide essentially the same results as conventional monitoring wells screened over the same interval (Butler et al. 2002, McCall et al., 2002).
Where did the slug come from?
Slug Test. No, it’s not a test to see if you have slimy worms crawling around in your garden! When single well hydraulic conductivity tests first got started, the wells being tested were pretty large in diameter ... 4-in., 8-in. or even larger. There were no neat small pressure transducers, no data loggers, and definitely no portable computers at this time. The tests were completed by pouring a 5-gallon bucket of water down the well casing and measuring the change in water level with a tape, marking time using your wrist watch, and writing down the time and water level in your field book with your No. 2 lead pencil! The 5-gallons of water was simply a big “slug” of water poured down the well. And presto! The “Slug Test” was born.